@article{oai:sucra.repo.nii.ac.jp:00017739, author = {落合, 陸 and 傳田, 晟矢 and 近藤, 一史 and 大向, 隆三}, issue = {1}, journal = {埼玉大学紀要. 教育学部, Journal of Saitama University. Faculty of Education}, month = {}, note = {We developed a subtraction circuit to measure the saturation spectra of rubidium atoms without using a lock-in amplifier. An instrumentation amplifier (IA), which is a type of differential amplifier, was used. We designed our IA to have an adjustable gain of 4.3 - 21 to obtain saturation spectra with a good signal-to-noise ratio. With our IA, we tried to deduce the saturation spectrum by subtracting the Doppler signal from the transmission spectrum of rubidium atoms enclosed in a glass cell. As a result, the obtained saturation spectra have six peaks originating from the atomic and cross-over resonances, and their line widths are 11.2 MHz. This number of peaks is the same as that detected with a lock-in amplifier, and the line width is smaller than that detected with a lock-in amplifier. Therefore, we confirmed that our IA has equivalent performance to a lock-in amplifier for measuring the saturation spectrum of atoms., text, application/pdf}, pages = {155--162}, title = {減算回路を用いた飽和吸収スペクトルの測定<数学・自然科学>}, volume = {64}, year = {2015}, yomi = {オチアイ, リク and デンダ, セイヤ and コンドウ, ヒトシ and オオムカイ, リュウゾウ} }